German make ceramic heater with parabolic reflector to maintain uniform temp across the mattress which swivels 90° on both side Inbuilt Observation Lamp.
Skin/Air and Manual Controller
Capable of holding up to 10 kg weight
Rectangular trolley with four sides clear acrylic panels. Front and both side collapsible, Easy Mattress Tilting in both directions
More than two infusion pumps can be clamped on Stainless Steel Rod IV Saline Stand Height can be adjusted
Middle Tray provision for bottom surface Phototherapy unit
Mounted on anti skid castors with locking facility
Display | : Baby / Air Temp Display Set Temp. Display Heater % Display |
Modes | : Baby, Air & Manual |
LED's for Alarms | : High Temp & Low Temp Heater Fail, Probe Fail, Power Fail |
Switchess | : On/Off Switch for warmer & observation light |
Display | : Baby Temp. Display Set Temp. Display Heater % Display |
Modes | : Baby & Manual |
LED's for Alarms | : High Temp & Low Temp Heater Fail, Probe Fail, Power Fail |
Switchess | : On/Off Switch for warmer & observation light |
Width : 80 cm
Depth : 85 cm
Height : 178 cm
Bed Size : 54x72 cm
Monitor Tray Size: 31 x26 cm
Mode : Skin/Air and Manual
Heater : Ceramic Heater - 650W
Heater Rotation : ±90° on both side
Reflector : Parabolic Reflector
Observation Lamp : LED 6W Examination Light
Monitor Tray Side Panels : Mild Steel Powder Coated
Side Panels : Unbreakable Clear Acrylic Panel
Wheels : Strong Castor Wheels with two brake
Mattress : Waterproof, Flame Resistance & Zipped Cover
Mattress Foam Density : 21-25 kg/m3
IV Pole : Adjustable Height
Infusion Pump Stand : Stainless Steel Rod
Temperature Sensor : Thermistor based NTE Sensors with high resolution
Timer : APGAR Timer
Supply Voltage : 190V-250V AC
Supply Frequency : 50-60 Hz
Power Consumption : 750W
Temperature Display : Celsius/Fahrenheit
Temperature Accuracy : ±0.2℃
Set Temperature Range : 30°-40℃
Heater Output : 0-100%
Modes of Operation : Baby, Air & Manual
Manual Mode Timer : 10 Min
Alarms : Audio with 1Omin Mute & Visual
High Temperature : If Temp > Set Temp + 1℃
Low Temperature : If Temp< Set Temp-1°C
Probe Failure : Probe Disconnection / Not Working
Power Failure : If Power Fails
Heater Failure : If Heater Stops working